• Dépistage, diagnostic, pronostic

  • Essais de technologies et de biomarqueurs dans un contexte clinique

  • Sein

Prospective Study of Supplemental Screening With Contrast-Enhanced Mammography in Women With Elevated Risk of Breast Cancer: Results of the Prevalence Round

Menée auprès de 460 femmes présentant un risque élevé de cancer du sein et dont les résultats de la tomosynthèse numérique 2D se sont avérés négatifs (âge minimum : 35 ans ; âge médian : 56,8 ans), cette étude évalue l'efficacité, du point de vue du taux de détection, d'une mammographie supplémentaire avec rehaussement de contraste

Purpose : Contrast-enhanced mammography (CEM) and magnetic resonance imaging (MRI) have shown similar diagnostic performance in detection of breast cancer. Limited CEM data are available for high-risk breast cancer screening. The purpose of the study was to prospectively investigate the efficacy of supplemental screening CEM in elevated risk patients.

Materials and Methods : A prospective, single-institution, institutional review board–approved observational study was conducted in asymptomatic elevated risk women age 35 years or older who had a negative conventional two-dimensional digital breast tomosynthesis screening mammography (MG) and no additional supplemental screening within the prior 12 months.

Results : Four hundred sixty women were enrolled from February 2019 to April 2021. The median age was 56.8 (range, 35.0-79.2) years; 408 of 460 (88.7%) were mammographically dense. Biopsy revealed benign changes in 22 women (22/37, 59%), high-risk lesions in four women (4/37, 11%), and breast cancer in 11 women (11/37, 30%). Fourteen cancers (10 invasive, tumor size range 4-15 mm, median 9 mm) were diagnosed in 11 women. The overall supplemental cancer detection rate was 23.9 per 1,000 patients, 95% CI (12.0 to 42.4). All cancers were grade 1 or 2, ER+ ERBB2–, and node negative. CEM imaging screening offered high specificity (0.875 [95% CI, 0.844 to 0.906]), high NPV (0.998 [95% CI, 0.993 to 1.000), moderate PPV1 (0.164 [95% CI, 0.076 to 0.253), moderate PPV3 (0.275 [95% CI, 0.137 to 0.413]), and high sensitivity (0.917 [95% CI, 0.760 to 1.000]). At least 1 year of imaging follow-up was available on all patients, and one interval cancer was detected on breast MRI 4 months after negative screening CEM.

Conclusion : A pilot trial demonstrates a supplemental cancer detection rate of 23.9 per 1,000 in women at an elevated risk for breast cancer. Larger, multi-institutional, multiyear CEM trials in patients at elevated risk are needed for validation.

Journal of Clinical Oncology , résumé, 2023